In September 2026, Mark, Quality Manager at CieloWorks, reviewed 23 FDA warning letters sent to medical device manufacturers that year. The top three problem areas were deficiencies with design controls, Corrective and Preventive Actions (CAPA), and complaint handling. Design control issues made up 18% of the inspection findings described in the warning letters reviewed.
Of the design control findings, half were specifically issues with design verification and validation.
Design Changes Without Adequate Validation Testing
In one example, a company made design changes that the design team had determined were insignificant and did not pose additional risk. The team therefore had not conducted validation testing to support that determination.
The FDA disagreed with the company’s decision.
This example highlights the importance of supporting design change decisions with adequate testing and documented justification. Reviewing these decisions before implementation can help manufacturers avoid additional testing, documentation rework, and delays.
Design Validation Issues Identified in FDA Warning Letters
The design validation findings reviewed included the following:
- Design validation did not include adequate testing of production units under actual or simulated use conditions.
- Design validation did not adequately ensure that the device conformed to all defined user needs.
- The design team failed to validate a product with an interoperable product and did not document a justification for not doing so.
- The company failed to adequately establish and maintain procedures for validating the device design.
- The design team determined that only design verification, not design validation, was required for a user need, but did not document a justification for that determination.
- A design change had been identified as not requiring design validation. The company’s initial response also acknowledged that it did not maintain documentation justifying the exclusion of worst-case physiological variables from its Human Factors Validation Report.
- Design validation procedures did not adequately ensure that devices conformed to defined user needs and intended uses, including testing of production units under actual or simulated use conditions.
- Procedures for validating the device design did not include an adequate risk analysis.
- The company had identified a hazard, but evidence of risk control implementation and validation was absent from the product’s risk analysis.
These findings demonstrate the need for medical device design validation procedures that address user needs, intended uses, applicable use conditions, and risk controls. Decisions about the scope of validation also need documented justification and supporting evidence.
Design Verification Issues Identified in FDA Warning Letters
The FDA also noted design verification issues. These included:
- Failure to confirm that design outputs met design input requirements during design verification.
- Failure to include predefined acceptance criteria for testing during design verification.
- Inadequate design verification testing despite complaint data that included approximately 5,000 complaints indicating that testing remained inadequate.
Clear acceptance criteria and documented results help demonstrate that design verification has addressed the specified requirements. Complaint data should also be evaluated for indications that verification testing or the underlying design requires further review.
Design Controls Under the FDA’s QMSR
The warning letter findings included references to the former design control requirements under 21 CFR 820.30. The FDA’s Quality Management System Regulation (QMSR), effective February 2, 2026, incorporates ISO 13485:2016 by reference.
Manufacturers should review their design and development procedures against the applicable current requirements, including those for design verification and validation.
How CieloWorks Can Help Prevent Design Control Issues
A review of verification and validation procedures can help identify gaps before they result in inspection findings or additional work. Addressing missing acceptance criteria, incomplete validation, or undocumented design change decisions earlier can reduce repeat testing, documentation rework, and project delays, helping manufacturers save time and money.