In September 2026, Mark, Quality Manager at CieloWorks, reviewed 23 FDA warning letters sent to medical device manufacturers that year. The top three problem areas were deficiencies with design controls, Corrective and Preventive Actions (CAPA), and complaint handling. Design control issues made up 18% of the inspection findings described in the warning letters reviewed.

Of the design control findings, half were specifically issues with design verification and validation.

Design Changes Without Adequate Validation Testing

In one example, a company made design changes that the design team had determined were insignificant and did not pose additional risk. The team therefore had not conducted validation testing to support that determination.

The FDA disagreed with the company’s decision.

This example highlights the importance of supporting design change decisions with adequate testing and documented justification. Reviewing these decisions before implementation can help manufacturers avoid additional testing, documentation rework, and delays.

Design Validation Issues Identified in FDA Warning Letters

The design validation findings reviewed included the following:

These findings demonstrate the need for medical device design validation procedures that address user needs, intended uses, applicable use conditions, and risk controls. Decisions about the scope of validation also need documented justification and supporting evidence.

Design Verification Issues Identified in FDA Warning Letters

The FDA also noted design verification issues. These included:

Clear acceptance criteria and documented results help demonstrate that design verification has addressed the specified requirements. Complaint data should also be evaluated for indications that verification testing or the underlying design requires further review.

Design Controls Under the FDA’s QMSR

The warning letter findings included references to the former design control requirements under 21 CFR 820.30. The FDA’s Quality Management System Regulation (QMSR), effective February 2, 2026, incorporates ISO 13485:2016 by reference.

Manufacturers should review their design and development procedures against the applicable current requirements, including those for design verification and validation.

How CieloWorks Can Help Prevent Design Control Issues

A review of verification and validation procedures can help identify gaps before they result in inspection findings or additional work. Addressing missing acceptance criteria, incomplete validation, or undocumented design change decisions earlier can reduce repeat testing, documentation rework, and project delays, helping manufacturers save time and money.

CieloWorks provides quality consulting and regulatory support, including reviews of medical device files and quality management systems. Its combined quality, regulatory, and medical software development expertise helps teams identify gaps and determine the work needed to address them.

If a company needs a review of its verification and validation procedures or design files in advance of an inspection, contact CieloWorks to discuss how an early review can help prevent these issues.